Anritsu MT8801 Communication Analyser

34 Anritsu MT8801B

Problem

Radio communications analyser performs several functions, including acting as a radio receiver and spectrum analyser. The spectrum analyser function was displaying the frequency spectrum shifted by a random and inconsistent frequency offset.

The spectrum analyser has a heterodyne structure with two stages of mixing. The first local oscillator is swept for tuning, provided by a phase locked loop (PLL) with a YIG oscillator.

The second mixing stage shifts the signal to a low intermediate frequency, using a phase locked loop with a dielectric resonance oscillator (DRO).

Both phase locked loops were becoming unlocked causing the frequency shifts that are seen on the spectrum analyser display.

35 Anritsu MT8801B

Diagnosis

Checking the many electrolytic power smoothing capacitors showed many were beginning to fail, the equivalent series resistance was higher than it should be.

All these needed to be replaced to prevent further issues appearing in future, and to reduce power supply noise.

Using a spectrum analyser and frequency counter to monitor the local oscillator behaviour, it was possible to check the stability of the signals while operating the analyser across different frequency ranges.

By examining the feedback to the phase detector in the PLL it became clear that an intermittent loss of signal was causing the unlocked condition. This led us to look for a poor connection.

38 Anritsu MT8801B

Solution

Replacing the many capacitors that were beginning to fail improved the stability of the PLLs but did not completely resolve the issues.

Adjusting the DRO plunger of the second local oscillator brought the startup frequency closer to the centre of the PLL capture range and allowed the PLL to remain locked correctly as the unit warmed up.

Part of the swept local oscillator PLL feedback loop was failing and the signal was disappearing. The cause was traced to a broken solder joint, losing bias current for a single transistor amplifier. With this fixed, the signal to the phase detector was reliable and the unit operates correctly.

If you are experiencing similar issues and would like to discuss or request a quote, email us at This email address is being protected from spambots. You need JavaScript enabled to view it.
or call 01303 489 698.

 

01 Anritsu MT8801B02 Anritsu MT8801B03 Anritsu MT8801B04 Anritsu MT8801B05 Anritsu MT8801B06 Anritsu MT8801B07 Anritsu MT8801B08 Anritsu MT8801B09 Anritsu MT8801B10 Anritsu MT8801B11 Anritsu MT8801B12 Anritsu MT8801B13 Anritsu MT8801B14 Anritsu MT8801B15 Anritsu MT8801B16 Anritsu MT8801B17 Anritsu MT8801B18 Anritsu MT8801B19 Anritsu MT8801B20 Anritsu MT8801B21 Anritsu MT8801B22 Anritsu MT8801B23 Anritsu MT8801B24 Anritsu MT8801B25 Anritsu MT8801B26 Anritsu MT8801B27 Anritsu MT8801B28 Anritsu MT8801B29 Anritsu MT8801B30 Anritsu MT8801B31 Anritsu MT8801B32 Anritsu MT8801B33 Anritsu MT8801B36 Anritsu MT8801B37 Anritsu MT8801B39 Anritsu MT8801B40 Anritsu MT8801B

 
 